Title: Evaluating the Impact Exposure Parameters Have on Automatic Tube Current Modulation in a CT Scanner
Authors: Hishaam, U.
Jeyasugiththan, J.
Satharasinghe, D. M.
Senanayake, G.
Kumarihami, A. M. C.
Ranaweera, G.
Costa, M.
Pallewatte, A. S.
Peterson, S.
Keywords: Computed Tomography
Patient Radiation
Automatic Tube Current Modulation
Dose Optimization
Issue Date: 2022
Publisher: Asia‑Oceania Federation of Organizations (AFOMP) Congress
Abstract: Background or Introduction: ATCM reduces dose by adapting tube current to maintain a specified target image quality throughout CT scans and across patients of varying sizes. Scanners modulate current differently: angularly around the patient or along the z-axis; using a reference image or chosen noise/dose reference levels. When misapplied, ATCM can subject patients to excessive radiation doses or produce images of non-diagnostic quality. Thus, evaluating the impact scanner parameters have on radiation dose is vital. Material and Methods: A 2-step phantom containing an air pocket, to mimic the lungs, was used to conduct a preliminary study(Philips BRILLIANCE iCT Scanner) where parameters were varied to evaluate their impact on tube current. Using the results, a customized phantom—5 diameter steps(16- 32cm), a cone, and an air pocket—was built. A comprehensive study on how varying pitch values, tube voltage, and DoseRight index(ATCM setting) impact the TCM for patients represented by this phantom followed. Results & Discussion: Preliminary Study: changing tube voltage from 120kVp-100kVp increased the tube current—patient radiation exposure—by 38%, and 42% when changed from 120kVp-80kVp. At 120kVP, when the DoseRight Index was changed from 15-20, current values increased by 42%, and from 15-25 this increase was 68%. At 100kVp, when DRI changed from 10-15 the current increased by about 43%, and 10-20 showed a 67% increase. Similarly, at 80kVp, a change of DRI from 10-15 resulted in a 33% increase in tube current. Customized phantom: Data to be collected in September. Conclusion(s): With ATCM, reducing the tube voltage results in an increase in radiation dose; to compensate for the associated increase in image noise. For multislice CT scanners, pitch value has minimal impact, but DoseRight Index shows a strong positive correlation with tube current. In-depth analysis of how changes to scanner parameters impact patients of varying sizes will be conducted when the study employing the customized phantom is completed.
URI: http://archive.cmb.ac.lk:8080/xmlui/handle/70130/6951
Appears in Collections:Department of Nuclear Science

Files in This Item:
File Description SizeFormat 
Abstract AFOMP 2022.pdf104.69 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.